HELIOS 3000 CEILING MOUNTED DENTAL LIGHT HL3C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-07-26 for HELIOS 3000 CEILING MOUNTED DENTAL LIGHT HL3C manufactured by Pelton & Crane.

Event Text Entries

[81260558] Upon evaluation by the distributor, it was determined the set screws were not properly secured by the distributor during installation. The set screws will prevent the light from unscrewing from the ceiling mount arm after installation. The pelton & crane installation instructions clearly states to secure all set screws during installation of the ceiling light. The installation instructions also lists warnings to ensure the set screws are properly tightened. The distributor service technician reattached the light and secured the sets screws.
Patient Sequence No: 1, Text Type: N, H10


[81260559] It was reported that a dr. Was positioning a pelton & crane dental light for use when the light fell down and onto the patients stomach. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2017-00043
MDR Report Key6743932
Date Received2017-07-26
Date of Report2017-06-30
Date of Event2017-06-29
Date Mfgr Received2017-06-30
Device Manufacturer Date2011-08-24
Date Added to Maude2017-07-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FRANK RAY
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1PELTON & CRANE
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHELIOS 3000 CEILING MOUNTED DENTAL LIGHT
Generic NameDENTAL LIGHT
Product CodeEAZ
Date Received2017-07-26
Model NumberHL3C
Catalog NumberHL3C
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-26