CHROMID? MRSA 43451

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-07-28 for CHROMID? MRSA 43451 manufactured by Biom?rieux Sa.

Event Text Entries

[81362444] A customer from (b)(6) reported to biom? Rieux a false positive (b)(6) result for a multi-resistant patient isolate in association with chromid? (b)(6) 20 plt. The customer reported the (b)(6) isolate was cultured on the chromid? (b)(6) and the isolate grew only the (b)(6) lot. The customer reported patient results were affected as infection control acts on presumptive positives that are green on chromid? Agar. The patient was isolated with other colonized (b)(6) positive patients due to this result and was possibly exposed to truly positive (b)(6) patients. The customer stated there was a delay in reporting results as sensitivity testing and further testing had to be performed due to the discrepancy. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00211
MDR Report Key6751003
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-07-28
Date of Report2017-11-21
Date Mfgr Received2017-10-31
Device Manufacturer Date2017-04-14
Date Added to Maude2017-07-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOM
Manufacturer Street5, RUE DES AQUEDUCS
Manufacturer CityCRAPONNE, FR 69290
Manufacturer CountryFR
Manufacturer Postal Code69290
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? MRSA
Generic NameCHROMID? MRSA
Product CodeJSO
Date Received2017-07-28
Catalog Number43451
Lot Number1005639900
ID Number03573026136642
Device Expiration Date2017-06-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer Address5, RUE DES AQUEDUCS CRAPONNE, FR 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-07-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.