HYDRO LEMAITRE VALVULOTOME 1009-00J

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-01 for HYDRO LEMAITRE VALVULOTOME 1009-00J manufactured by Lemaitre Vascular, Inc..

Event Text Entries

[81728676] We have received and evaluated the complaint device. We were able to confirm the reported incident. We found that one of the blades did not insert into the retainer intermittently when the centering hoops were opened and closed multiple times. We also confirmed that it was the same blade that did not insert into the retainer. Our lot history records review did not reveal any discrepancies related to the complaint event either in the manufacturing or packaging processes. Please note that we do conduct 100% inspection of the blade assembly during the manufacturing process. Our quality control also samples these devices before final packaging to ensure proper blade adjustment. It is possible that the device was damaged during the manufacturing process or after inspection and prior to packaging. As a result, we have implemented corrective action to minimize the possibility of this recurring.
Patient Sequence No: 1, Text Type: N, H10


[81728677] During pre-use check, one of the blade was unable to close completely into the retainer.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1220948-2017-00044
MDR Report Key6758093
Date Received2017-08-01
Date of Report2017-08-01
Date of Event2017-07-03
Date Mfgr Received2017-07-03
Device Manufacturer Date2016-11-17
Date Added to Maude2017-08-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. PRAGYA THIKEY
Manufacturer Street63 SECOND AVE
Manufacturer CityBURLINGTON MA 01803
Manufacturer CountryUS
Manufacturer Postal01803
Manufacturer Phone7812212266
Manufacturer G1LEMAITRE VASCULAR, INC.
Manufacturer Street63 SECOND AVE
Manufacturer CityBURLINGTON MA 01803
Manufacturer CountryUS
Manufacturer Postal Code01803
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHYDRO LEMAITRE VALVULOTOME
Generic NameVALVULOTOME
Product CodeMGZ
Date Received2017-08-01
Returned To Mfg2017-07-25
Catalog Number1009-00J
Lot NumberELVH1158V
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEMAITRE VASCULAR, INC.
Manufacturer Address63 SECOND AVE BURLINGTON MA 01803 US 01803


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.