ARCHITECT C4000 ANALYZER 6L45-21 02P24-40

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-08-03 for ARCHITECT C4000 ANALYZER 6L45-21 02P24-40 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[81813757] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[81813758] The customer stated that an architect total bilirubin result of 17. 2 mg/dl was reported for a newborn patient on (b)(6) 2017. Repeat testing (per reflex rule set up by the customer) was 7. 4 mg/dl. The laboratory technician released the result of 17. 2 mg/dl without checking for the reflex result. The baby was sent to the nicu where retesting for total bilirubin was 7. 6 mg/dl. The customer is not aware of any treatment given to the baby. The repeat testing in nicu prompted the lookback at the architect results where the reflex result was discovered. No known unnecessary treatment was given and no injury was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2017-00301
MDR Report Key6764215
Report SourceHEALTH PROFESSIONAL
Date Received2017-08-03
Date of Report2018-07-31
Date of Event2017-07-11
Date Mfgr Received2018-07-02
Date Added to Maude2017-08-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Removal Correction Number1628664-07/17/18-001-C
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameARCHITECT C4000 ANALYZER
Generic NameAUTOMATED CHEMISTRY ANALYZER
Product CodeMQM
Date Received2017-08-03
Model Number6L45-21
Catalog Number02P24-40
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038

Device Sequence Number: 1

Brand NameARCHITECT C4000 ANALYZER
Generic NameAUTOMATED CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-08-03
Model Number6L45-21
Catalog Number02P24-40
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.