COBAS E 411 IMMUNOASSAY ANALYZER E411 RACK 04775201001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-08-04 for COBAS E 411 IMMUNOASSAY ANALYZER E411 RACK 04775201001 manufactured by Roche Diagnostics.

Event Text Entries

[82581317] (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[82581318] The customer complained of a low erroneous result for 1 patient sample tested for elecsys prolactin assay (prolactin) on a cobas e 411 immunoassay analyzer. The initial result was 0. 076 ng/ml. This result was reported outside of the laboratory. The sample was repeated and the result was 15. 55 ng/ml. This result was also reported outside of the laboratory. Quality control (qc) results were within the acceptable range. There was no allegation that an adverse event occurred. The prolactin reagent lot number was 17223700 with an expiration date of 03/31/2018. The customer noted that there was a sample noise alarm at the same time this sample was pipetted and is wondering why there was no instrument alarm produced with the low result. The customer also stated there was wet serum on the wall of the sample container.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01602
MDR Report Key6767508
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-08-04
Date of Report2017-09-12
Date of Event2017-07-12
Date Mfgr Received2017-07-14
Date Added to Maude2017-08-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS E 411 IMMUNOASSAY ANALYZER
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeCFT
Date Received2017-08-04
Model NumberE411 RACK
Catalog Number04775201001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS E 411 IMMUNOASSAY ANALYZER
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-08-04
Model NumberE411 RACK
Catalog Number04775201001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.