COMPACT 300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-04 for COMPACT 300 manufactured by Ge Medical Systems, Llc.

Event Text Entries

[82011234] Ge healthcare's investigation is ongoing. A follow-up report will be submitted once the investigation has been completed. Due to country specific privacy laws, patient information is not available. Device evaluation anticipated, but not yet begun.
Patient Sequence No: 1, Text Type: N, H10


[82011235] On (b)(6) 2017, the technologist at (b)(6) hospital in (b)(6) stated that during a wall stand chest radiographic exam using the compact 300 system, the collimator detached from the tube column. The collimator detached as the technologist was angulating for patient positioning. The collimator fell and struck the table and then fell to the floor. The collimator did not impact any person so there was no injury related to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2126677-2017-00010
MDR Report Key6767920
Date Received2017-08-04
Date of Report2017-10-18
Date Mfgr Received2017-10-12
Device Manufacturer Date1970-01-01
Date Added to Maude2017-08-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNE KRUEGER
Manufacturer Street3000 NORTH GRANDVIEW BOULEVARD
Manufacturer CityWAUKESHA WI
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOMPACT 300
Generic NameXR COMPACT300-MPS
Product CodeIZO
Date Received2017-08-04
ID NumberUDI NOT REQUIRED
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGE MEDICAL SYSTEMS, LLC
Manufacturer Address3000 NORTH GRANDVIEW BOULEVARD WAUKESHA, WI US


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-04