COBAS E 411 IMMUNOASSAY ANALYZER E411 DISK

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-08-07 for COBAS E 411 IMMUNOASSAY ANALYZER E411 DISK manufactured by Roche Diagnostics.

Event Text Entries

[83311720] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[83311721] The customer received a questionable low elecsys afp assay result for one patient sample. The initial result was 4. 00 iu/ml and was reported outside the laboratory. The patient had a sacrococcygeal soft tissue mass between gluteal muscles (yolk sac tumor). After an operation, the mass was determined to be malignant and the laboratory was informed. The same sample was repeated 10 times on (b)(6) 2017 and the result was consistently >1000 iu/ml with a data flag. There was no allegation of an adverse event. The reagent lot number was 17674700 with an expiration date of 10/31/2017. The current date set on the analyzer was backdated for an unknown reason. Review of the provided calibration data did not find any issues and confirmed it was acceptable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01621
MDR Report Key6770153
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-08-07
Date of Report2017-09-06
Date of Event2017-06-25
Date Mfgr Received2017-07-19
Date Added to Maude2017-08-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS E 411 IMMUNOASSAY ANALYZER
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeCDZ
Date Received2017-08-07
Model NumberE411 DISK
Catalog NumberASKU
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS E 411 IMMUNOASSAY ANALYZER
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-08-07
Model NumberE411 DISK
Catalog NumberASKU
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-07

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.