MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-08-07 for COBAS E 411 IMMUNOASSAY ANALYZER E411 DISK manufactured by Roche Diagnostics.
[83311720]
(b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[83311721]
The customer received a questionable low elecsys afp assay result for one patient sample. The initial result was 4. 00 iu/ml and was reported outside the laboratory. The patient had a sacrococcygeal soft tissue mass between gluteal muscles (yolk sac tumor). After an operation, the mass was determined to be malignant and the laboratory was informed. The same sample was repeated 10 times on (b)(6) 2017 and the result was consistently >1000 iu/ml with a data flag. There was no allegation of an adverse event. The reagent lot number was 17674700 with an expiration date of 10/31/2017. The current date set on the analyzer was backdated for an unknown reason. Review of the provided calibration data did not find any issues and confirmed it was acceptable.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2017-01621 |
MDR Report Key | 6770153 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2017-08-07 |
Date of Report | 2017-09-06 |
Date of Event | 2017-06-25 |
Date Mfgr Received | 2017-07-19 |
Date Added to Maude | 2017-08-07 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | HITACHI HIGH TECH CORP. |
Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
Manufacturer City | IBARAKI 312-8504 |
Manufacturer Country | JA |
Manufacturer Postal Code | 312-8504 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS E 411 IMMUNOASSAY ANALYZER |
Generic Name | IMMUNOCHEMISTRY ANALYZER |
Product Code | CDZ |
Date Received | 2017-08-07 |
Model Number | E411 DISK |
Catalog Number | ASKU |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Brand Name | COBAS E 411 IMMUNOASSAY ANALYZER |
Generic Name | IMMUNOCHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2017-08-07 |
Model Number | E411 DISK |
Catalog Number | ASKU |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-08-07 |