LUXSTAR TRACK-MOUNT DENTAL LIGHT TL1000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-08-08 for LUXSTAR TRACK-MOUNT DENTAL LIGHT TL1000 manufactured by Marus.

Event Text Entries

[82106979] The lens heat shield cover from the dental light has a three prong locking mechanism which locks the lens heat shield cover into place. The lens heat shield cover has to be removed by the end user when cleaning the lens or replacing the halogen bulb during routine operator maintenance. The installation instructions, use and care instructions, and labeling clearly state to ensure the lens heat shield cover is properly secure by snapping the part back into place. Marus initiated a recall on june 10, 2011 to add a tether to the lens heat shield cover to keep it from falling off the dental light if it is not re-installed properly. The fda closed this recall on 10/26/2012 (please reference recall z-2834-2011). Marus notified all distributors of the recall while the recall was open. The distributor responded back to marus and acknowledged the recall but apparently did not install the tether kit at this one particular dental practice. After following back up with the dentist on (b)(6) 2017, the dentist informed marus the tether was installed on the dental light and is working properly.
Patient Sequence No: 1, Text Type: N, H10


[82106980] A dentist was positioning the dental light for use when the lens heat shield cover fell off the dental light and onto the patients neck causing a burn. The dr. Applied over the counter burn cream to the affected area. The patient followed up with the dentist 3 weeks later and the dentist believes the burn may be turning into a scar.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2017-00046
MDR Report Key6774856
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-08-08
Date of Report2017-07-20
Date of Event2017-06-23
Date Mfgr Received2017-07-20
Device Manufacturer Date2010-10-22
Date Added to Maude2017-08-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FRANK RAY
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1MARUS
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLUXSTAR TRACK-MOUNT DENTAL LIGHT
Generic NameDENTAL LIGHT
Product CodeEAZ
Date Received2017-08-08
Model NumberTL1000
Catalog NumberTL1000
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMARUS
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-08-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.