OLYMPIC 56320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-07 for OLYMPIC 56320 manufactured by Olympic Now Known As Natus Medical.

Event Text Entries

[82370042] Olympic neonatal scale plugged into outlet and plug blew and nurse received mild shock. Investigation of equipment by bio-med engineering revealed the cord cap on the scale was damaged. The cord cap was replaced and the scale was checked and returned to service. Hospital electrician also checked all the outlets in the room and they were found to be in good working order.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5071443
MDR Report Key6776677
Date Received2017-08-07
Date of Report2017-08-07
Date of Event2017-07-24
Date Added to Maude2017-08-09
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 0

Brand NameOLYMPIC
Generic NameNEONATAL SCALE
Product CodeFRW
Date Received2017-08-07
Model Number56320
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrI
Device Sequence No0
Device Event Key0
ManufacturerOLYMPIC NOW KNOWN AS NATUS MEDICAL

Device Sequence Number: 1

Brand NameOLYMPIC
Generic NameNEONATAL SCALE
Product CodeFRW
Date Received2017-08-07
Model Number56320
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS NOW KNOWN AS NATUS MEDICAL


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-08-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.