PATHROMTIN SL 10484200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-08-11 for PATHROMTIN SL 10484200 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[82668018] Siemens healthcare diagnostics personnel reviewed the backup files provided by the customer of the bcs xp analyzer to determine the cause of the high activated partial prothrombin time (aptt) result. Review of the bcs xp analyzer backup file concluded that no failure occurred during measurement of the sample. A difference between the results obtained on the bcs xp system and the cs-5100 system may be due to the differences in method application used and differences in the clot detection algorithms. The instrument and reagent are performing according to specifications. No further evaluation of this device is required. Mdr 9610806-2017-00088 was filed for the same event.
Patient Sequence No: 1, Text Type: N, H10


[82668019] A high activated partial thromboplastin time (aptt) result of 45. 0sec on one patient sample was generated on the bcs xp analyzer using the pathromtin sl assay. The same sample was run on the cs-5100 analyzer using the pathromtin sl assay and an aptt result of "no clot" was generated. Neither result was reported to the physician. It is unknown what the expected aptt result should have been for the patient. All qc recovery was acceptable. There are no reports of patient intervention or adverse health consequence due to the high aptt result of 45. 0sec as the result was not reported to the physician.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2017-00089
MDR Report Key6789797
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-08-11
Date of Report2017-11-15
Date of Event2017-07-19
Date Mfgr Received2017-11-15
Device Manufacturer Date2017-01-18
Date Added to Maude2017-08-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLORIANN RUSSO
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242287
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer StreetEMIL VON BEHRING STRASSE 76
Manufacturer CityMARBURG, 35041
Manufacturer CountryGM
Manufacturer Postal Code35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePATHROMTIN SL
Generic NamePATHROMTIN SL
Product CodeGFO
Date Received2017-08-11
Catalog Number10484200
Lot Number536695
Device Expiration Date2018-07-24
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL VON BEHRING STRASSE 76 MARBURG, 35041 GM 35041


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.