80-190 ML FILL VOLUME CRYOGENIC STORAGE CONTAINER CF-750

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-14 for 80-190 ML FILL VOLUME CRYOGENIC STORAGE CONTAINER CF-750 manufactured by Charter Medical, Ltd..

Event Text Entries

[82937277] Four unused bags and three used bags were returned for investigation. No issues were observed on the unused bags. The bags were not used in a clinical setting. Irregular film fractures were observed on each of the three used bags. The film fractures were located on one side of all three returned used bags. The location and nature of the fractures suggested air bubbles may have been introduced into the bag when filling with fluid. The user provided pictures of cf-750 bags with fluid prior to freezing. Large bubbles were observed inside the bags. The bubbles inside the bags would correspond to the approximate area where the bag fractures were located on the three returned bags. There are other causes that could potentially result in bag fractures: - overfilling the bag. An overfilled bag could create resistance during insertion and removal of the bag from the metal cassette. - moisture on the exterior surface of the bag or cassette interior when the bag is inserted into the metal cassette. Moisture may cause the bag film to freeze to the interior cassette surface, thus potentially resulting in damage when the frozen bag is removed from the cassette. - an unknown material anomaly. The product instruction for use sheet is provided with each cf-750 device. Review of the ifu version provided with lot 149522 has the following precautions: - "port tubes and film are fragile in the frozen state and breakage may occur. Handle with care. " - "do not overfill. " - "remove as much air as possible from the container. " - "ensure bag exterior and protective freezing cassettes are dry prior to initiating freezing protocol. Moisture on exterior of bag or on cassette could cause adherence of bag to cassette resulting in difficulty of bag removal. "
Patient Sequence No: 1, Text Type: N, H10


[82937278] The bags were placed in ln2 on (b)(6). The cracks were discovered when the bags were removed from the ln2 tank prior to thawing on (b)(6). There was no contact with a patient following discovery of the bag breaks. There was no injury, death or further medical intervention required as a result of this occurrence. There was no medical staff exposure as a result of this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2017-00008
MDR Report Key6791907
Date Received2017-08-14
Date of Report2017-08-14
Date of Event2017-07-06
Date Mfgr Received2017-07-12
Device Manufacturer Date2017-05-23
Date Added to Maude2017-08-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA HUGHES
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name80-190 ML FILL VOLUME CRYOGENIC STORAGE CONTAINER
Generic Name80-190 ML FILL VOLUME CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2017-08-14
Returned To Mfg2017-07-14
Model NumberCF-750
Catalog NumberCF-750
Lot Number149522
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-14

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.