STERRAD? CHEMICAL INDICATOR STRIP 14100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-08-16 for STERRAD? CHEMICAL INDICATOR STRIP 14100 manufactured by Advanced Sterilization Products.

Event Text Entries

[82791567] (b)(4). (b)(6). Asp investigation summary: the investigation included a review of the device history record (dhr), trending of lot number, retains testing, concomitant product evaluation, visual analysis and system risk analysis (sra). ? The dhr was reviewed and the involved lot met manufacturer specifications at the time of release. No anomalies were observed that would contribute to the customer's experienced issue. ? Trending analysis by lot number was reviewed from 11/22/2016 to 5/22/2017 and no significant trend is observed. ? Thirty retains ci stips were subject to functional evaluation. All thirty ci strips met specification for color change post-sterrad? Processing. ? Concomitant product evaluation determined it is unlikely that the sterrad? 100s unit contributed to the issue since the cycle completed and the remainder of the ci strip changed color correctly. ? The product was not returned; however, visual analysis was performed from a photograph of the affected product provided by the customer. The picture confirms there is a "red spot" on the strip and the remainder of the strip has changed color correctly. It could not be determined if the red spot was a result of the ci strip being blocked during sterilization and the customer was unable to confirm if that was the situation or not. ? The sra indicates the risk associated with a exposure to biohazardous, pathogenic or infectious material is "low. " there is insufficient information to determine an assignable cause. It is unlikely that the color-chemical indicator issue was caused by a manufacturing issue since the retains product met functional specification, dhr review found no anomalies that would contribute to a incorrect color change and no significant trends were found in lot trending review. An issue with sterrad? Performance is also unlikely since the cycle passed and the remainder of the ci strip changed color correctly. The chemical indicator retains met functional specification. As a result, there is not an issue with product performance when used per the instructions for use (ifu). The issue will continue to be tracked and trended.
Patient Sequence No: 1, Text Type: N, H10


[82791568] A customer reported a sterrad? Chemical indicator strip did not change color correctly after a completed sterrad? 100s cycle. The affected load was released for use on patients. There was no report of infection, injury or harm to patient(s) associated with this issue. Although there is no report of patient injury or harm and no prior incidents have resulted in serious injury, advanced sterilization products (asp) has determined in this situation sterility cannot be assured. Therefore, as a matter of policy asp had decided to report all incidents of sterrad? Chemical indicator strips not changing color correctly and the load is released without reprocessing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2084725-2017-00535
MDR Report Key6797480
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-08-16
Date of Report2017-06-27
Date of Event2017-05-17
Date Mfgr Received2017-06-27
Device Manufacturer Date2016-08-08
Date Added to Maude2017-08-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JOAQUIN KURZ
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE
Manufacturer Phone949789-383
Manufacturer G1ADVANCED STERILIZATION PRODUCTS
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTERRAD? CHEMICAL INDICATOR STRIP
Generic NameCHEMICAL INDICATOR
Product CodeJOJ
Date Received2017-08-16
Catalog Number14100
Lot Number221611-02
Device Expiration Date2018-02-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerADVANCED STERILIZATION PRODUCTS
Manufacturer Address33 TECHNOLOGY DRIVE IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.