MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-08-16 for COBAS 6000 C (501) MODULE C501 04745914001 manufactured by Roche Diagnostics.
[83304558]
This event occurred in (b)(6). Unique identifier (udi)#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[83304559]
The customer complained of intermittent issues with low albumin results on a cobas 6000 c (501) module. The customer provided erroneous results for 1 patient. The initial albumin result was <3 g/l. This result was reported outside of the laboratory. The sample was repeated and the result was 39 g/l. There was no allegation that an adverse event occurred. The albumin reagent lot number and expiration date were not provided. The sample probe had been changed on (b)(6) 2017. The field service engineer (fse) visited the customer site on (b)(6) 2017 and changed the sample probe again. Quality control results were acceptable. The albumin results have been fine since the service visit. The investigation is ongoing.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2017-01715 |
MDR Report Key | 6798482 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2017-08-16 |
Date of Report | 2017-08-16 |
Date of Event | 2017-07-03 |
Date Mfgr Received | 2017-07-27 |
Date Added to Maude | 2017-08-16 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | HITACHI HIGH TECH CORP. |
Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
Manufacturer City | IBARAKI 312-8504 |
Manufacturer Country | JA |
Manufacturer Postal Code | 312-8504 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS 6000 C (501) MODULE |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | CJW |
Date Received | 2017-08-16 |
Model Number | C501 |
Catalog Number | 04745914001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Brand Name | COBAS 6000 C (501) MODULE |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2017-08-16 |
Model Number | C501 |
Catalog Number | 04745914001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-08-16 |