COBAS 6000 C (501) MODULE C501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-08-17 for COBAS 6000 C (501) MODULE C501 manufactured by Roche Diagnostics.

Event Text Entries

[83204577] This event occurred in (b)(6). Unique identifier (udi)#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[83204578] The customer complained of erroneous results for 1 patient sample tested for crpl3 c-reactive protein gen. 3 on a cobas 6000 c (501) module. The initial crpl3 result was 0. 30 mg/l. This result was reported outside of the laboratory where it was questioned by the doctor. The sample was repeated on a different c501 module and the result was 192. 42 mg/l. There was no allegation that an adverse event occurred. The crpl3 reagent lot number was 23811301. The expiration date was not provided. No issues were identified during a review of the calibration trace or the alarm trace provided by the customer. Based on the data available for investigation, it is clearly visible that no sample was pipetted. The customer had noted that the sample was quite turbid. The most likely root cause of the low result was due to the sample itself.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01721
MDR Report Key6799910
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-08-17
Date of Report2017-09-13
Date of Event2017-07-27
Date Mfgr Received2017-07-27
Date Added to Maude2017-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 6000 C (501) MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeDCN
Date Received2017-08-17
Model NumberC501
Catalog NumberASKU
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 6000 C (501) MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-08-17
Model NumberC501
Catalog NumberASKU
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-17

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.