MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-08-17 for COBAS 6000 C (501) MODULE C501 manufactured by Roche Diagnostics.
[83204577]
This event occurred in (b)(6). Unique identifier (udi)#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[83204578]
The customer complained of erroneous results for 1 patient sample tested for crpl3 c-reactive protein gen. 3 on a cobas 6000 c (501) module. The initial crpl3 result was 0. 30 mg/l. This result was reported outside of the laboratory where it was questioned by the doctor. The sample was repeated on a different c501 module and the result was 192. 42 mg/l. There was no allegation that an adverse event occurred. The crpl3 reagent lot number was 23811301. The expiration date was not provided. No issues were identified during a review of the calibration trace or the alarm trace provided by the customer. Based on the data available for investigation, it is clearly visible that no sample was pipetted. The customer had noted that the sample was quite turbid. The most likely root cause of the low result was due to the sample itself.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2017-01721 |
MDR Report Key | 6799910 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2017-08-17 |
Date of Report | 2017-09-13 |
Date of Event | 2017-07-27 |
Date Mfgr Received | 2017-07-27 |
Date Added to Maude | 2017-08-17 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | HITACHI HIGH TECH CORP. |
Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
Manufacturer City | IBARAKI 312-8504 |
Manufacturer Country | JA |
Manufacturer Postal Code | 312-8504 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS 6000 C (501) MODULE |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | DCN |
Date Received | 2017-08-17 |
Model Number | C501 |
Catalog Number | ASKU |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Brand Name | COBAS 6000 C (501) MODULE |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2017-08-17 |
Model Number | C501 |
Catalog Number | ASKU |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-08-17 |