ELECSYS HSV-1 IGG IMMUNOASSAY 05572185160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-17 for ELECSYS HSV-1 IGG IMMUNOASSAY 05572185160 manufactured by Roche Diagnostics.

Event Text Entries

[83749647] Unique identifier (udi)#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[83749648] The customer received a questionable elecsys (b)(6) igg immunoassay result for one patient sample. The customer used cobas 6000 e 601 module serial number (b)(4). The initial result was 0. 560 coi ((b)(6)) and was reported outside the laboratory. The doctor doubted the result as the patient was previously (b)(6). The sample was sent to a reference laboratory and tested on a siemens immulyte (both (b)(6)) and the result was (b)(6). No specific data was provided. The sample was then sent another reference laboratory and the result was 1. 94 coi ((b)(6)). The patient was redrawn on (b)(6) 2017 and the result was from the cobas e601 was again (b)(6). They sent the sample to another laboratory and the result was (b)(6). Specific data was not provided. The patient was not adversely affected. For troubleshooting, the customer repeated a proficiency sample and the results were acceptable. The field service representative could not find a cause. He confirmed the customer repeated the sample with calibration, qc, and proficiency material and all results were within specification. The field application specialist could not find a cause. He reviewed and verified all data including calibration, qc, and methodologies. He verified the assay was performing to specification.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01724
MDR Report Key6799984
Date Received2017-08-17
Date of Report2018-01-08
Date of Event2017-07-13
Date Mfgr Received2017-07-27
Date Added to Maude2017-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameELECSYS HSV-1 IGG IMMUNOASSAY
Generic NameHERPES SIMPLEX VIRUS, HSV-1
Product CodeMXJ
Date Received2017-08-17
Model NumberNA
Catalog Number05572185160
Lot Number214509
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.