COBAS E 411 IMMUNOASSAY ANALYZER E411 DISK 4775279001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-08-17 for COBAS E 411 IMMUNOASSAY ANALYZER E411 DISK 4775279001 manufactured by Roche Diagnostics.

Event Text Entries

[83750603] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[83750604] The customer received questionable elecsys total psa immunoassay and/or elecsys free psa immunoassay result for two patient samples. Of the data provided, only the following results were discrepant. Patient 1 initial total psa result was >100 ng/ml. The repeat result with a 1:5 dilution was 48. 76 ng/ml. The repeat result with no dilution was >100 ng/ml and the repeat result with a 1:2 dilution was 140 ng/ml. Patient 2 initial free psa result was 0. 003 ng/ml. The repeat result was 0. 5 ng/ml. This patient was male. No erroneous result was reported outside the laboratory. There was no allegation of an adverse event. The total psa reagent lot number was 19942201 with an expiration date of 04/30/2018. The free psa reagent lot number was 15099000 with an expiration date of 08/31/2017. The field service representative adjusted the mixer speed because he found it was creating foam in the reagents and he adjusted the position of the probe.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01726
MDR Report Key6800054
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-08-17
Date of Report2017-08-17
Date of Event2017-07-27
Date Mfgr Received2017-07-27
Date Added to Maude2017-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS E 411 IMMUNOASSAY ANALYZER
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeLTJ
Date Received2017-08-17
Model NumberE411 DISK
Catalog Number4775279001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS E 411 IMMUNOASSAY ANALYZER
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-08-17
Model NumberE411 DISK
Catalog Number4775279001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-17