ELEVANCE L/R LED DENTAL LIGHT 153901-002

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-08-17 for ELEVANCE L/R LED DENTAL LIGHT 153901-002 manufactured by Midmark Corporation.

Event Text Entries

[82957408] After review of pictures of the incident provided by the user facility the following explanation is the most likely cause of the light falling: the upper l/r light arm was out at a 90 degree angle from the dental chair toward a half-wall that was within the range of the light arm. When the dental chair was being lowered while the light was in this position, the light assembly would get caught on the half-wall which would cause the upper l/r arm to raise out of the joint. This would allow the light assembly and upper l/r arm to separate from lower l/r arm and tip toward the patient sitting in the dental chair. Additionally, the user facility stated they would not send the unit back for evaluation as they decided to assemble the dental light back together and resume using it. No servicing or replacements components were requested by the user facility.
Patient Sequence No: 1, Text Type: N, H10


[82957409] During a root canal procedure, the l/r led dental light fell on the head of a patient, cutting the patient's head. The doctor present at that time stitched up the laceration.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1523530-2017-00003
MDR Report Key6801679
Report SourceHEALTH PROFESSIONAL
Date Received2017-08-17
Date of Report2017-08-08
Date of Event2017-07-08
Date Mfgr Received2017-07-18
Device Manufacturer Date2014-03-20
Date Added to Maude2017-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ADAM CLUTTER
Manufacturer Street60 VISTA DRIVE
Manufacturer CityVERSAILLES OH 45380
Manufacturer CountryUS
Manufacturer Postal45380
Manufacturer Phone9375268474
Manufacturer G1MIDMARK CORPORATION
Manufacturer Street60 VISTA DRIVE
Manufacturer CityVERSAILLES OH 45380
Manufacturer CountryUS
Manufacturer Postal Code45380
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELEVANCE L/R LED DENTAL LIGHT
Generic NameDENTAL LED LIGHT
Product CodeEAZ
Date Received2017-08-17
Model Number153901-002
Catalog Number153901-002
OperatorDENTIST
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMIDMARK CORPORATION
Manufacturer Address60 VISTA DRIVE VERSAILLES OH 45380 US 45380


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-08-17

© 2026 FDA.report