FENIX CONTINENCE RESTORATION SYSTEM FXS16 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-08-17 for FENIX CONTINENCE RESTORATION SYSTEM FXS16 NA manufactured by Torax Medical, Inc..

Event Text Entries

[82977162] Following a surgical procedure for reinforcement of the anal sphincter due to fecal incontinence, a patient was found to have fenix beads eroded through the anal canal leading to fenix device explant. The fenix device was used as part of the surgical procedure. -uneventful surgical procedure and device implant on (b)(6)2017; 3-4 vessels were stitched during the implant procedure to decrease bleeding. -the patient presented with discharge and tenderness; during a digital rectal exam the device was found to have eroded into the anterior rectum on (b)(6)2017. Symptoms began (b)(6)2017. -uneventful device explant (b)(6)2017 due to 2 fenix device beads found eroded through the anal canal; an infection was present at the exposure site. The fenix device was removed transanally. The device was intact at the time of explant.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008766073-2017-00095
MDR Report Key6802365
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2017-08-17
Date of Report2017-08-31
Date of Event2017-07-20
Date Mfgr Received2017-08-31
Device Manufacturer Date2016-03-16
Date Added to Maude2017-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA AHLBORN
Manufacturer Street4188 LEXINGTON AVENUE NORTH
Manufacturer CitySHOREVIEW MN 55126
Manufacturer CountryUS
Manufacturer Postal55126
Manufacturer Phone6513618900
Manufacturer G1TORAX MEDICAL, INC.
Manufacturer Street4188 LEXINGTON AVENUE NORTH
Manufacturer CitySHOREVIEW MN 55126
Manufacturer CountryUS
Manufacturer Postal Code55126
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFENIX CONTINENCE RESTORATION SYSTEM
Generic NameIMPLANTED FECAL INCONTINENCE DEVICE
Product CodePMH
Date Received2017-08-17
Returned To Mfg2017-08-28
Model NumberFXS16
Catalog NumberNA
Lot Number10522
OperatorPHYSICIAN
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTORAX MEDICAL, INC.
Manufacturer Address4188 LEXINGTON AVENUE N SHOREVIEW MN 55126 US 55126


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Deathisabilit 2017-08-17

© 2026 FDA.report