MST MALYUGIN RING SYSTEM MAL-1002

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-08-17 for MST MALYUGIN RING SYSTEM MAL-1002 manufactured by Microsurgical Technology Inc.

Event Text Entries

[82976936] The doctor noticed the capsule tear towards the end of the procedure after the iol had been implanted. It is unclear why the malyugin ring was implicated as the cause. The device evaluation showed that there were no burrs or sharp edges present on the device. This complaint cannot be confirmed.
Patient Sequence No: 1, Text Type: N, H10


[82976937] The sales rep reported that the surgeon noticed a capsule tear after inserting an iol. A sulcus lens was used and the procedure was completed as planned.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019924-2017-00010
MDR Report Key6802951
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2017-08-17
Date of Report2017-08-16
Date of Event2017-07-11
Date Mfgr Received2017-07-19
Device Manufacturer Date2017-03-29
Date Added to Maude2017-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROB RANEY
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer Phone4255560544
Manufacturer G1MICROSURGICAL TECHNOLOGY INC
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal Code98052
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMST MALYUGIN RING SYSTEM
Generic NameIRIS RETRACTOR
Product CodeHOC
Date Received2017-08-17
Returned To Mfg2017-07-20
Model NumberMAL-1002
Catalog NumberMAL-1002
Lot Number087216
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY INC
Manufacturer Address8415 154TH AVE NE REDMOND WA 98052 US 98052


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-08-17

© 2026 FDA.report