COBAS 6000 E 601 MODULE E601 04745922001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-08-18 for COBAS 6000 E 601 MODULE E601 04745922001 manufactured by Roche Diagnostics.

Event Text Entries

[83749868] Unique identifier (udi)#: (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[83749869] The customer received questionable low elecsys total psa immunoassay and free psa results for one patient sample. The initial total psa result was 0. 681 ug/l and free psa result was 0. 311 ug/l. The same sample was repeated on (b)(6) 2017. The total psa result was 9. 91 ug/l with a data flag and the free psa result was 0. 600 ug/l. A new sample was drawn from the patient and the total psa result was 10. 59 ug/l and free psa result was 0. 671 ug/l. The erroneous result was not reported outside the laboratory. There was no allegation of an adverse event. The total psa reagent lot number was 19942201 and free psa reagent lot number was 209877. The expiration dates were requested but were not provided. A specific root cause could not be determined. A typical cause for low result is a sample pipetting issue. Review of the provided analyzer alarm trace found a few sample pipetting alarms. The field service representative checked the analyzer and preanalytics.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01745
MDR Report Key6803075
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-08-18
Date of Report2017-10-09
Date of Event2017-06-29
Date Mfgr Received2017-07-30
Date Added to Maude2017-08-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 6000 E 601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeLTJ
Date Received2017-08-18
Model NumberE601
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 6000 E 601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-08-18
Model NumberE601
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.