COVERED CP STENT 427 CVRDCP8Z45

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-18 for COVERED CP STENT 427 CVRDCP8Z45 manufactured by Numed, Inc..

Event Text Entries

[83279829] The covering detachment was confirmed. Four adhesive spots are visible on the stent. One end of the stent has been crimped down. The root cause of this complaint cannot be determined. All covered stents are inspected in final qc to make sure that each covering has (4) acceptable glue spots. The stents are visually inspected once more after they are packaged for covering adhesion. A sample from each lot is tested for covering strength. The sample from this lot failed at 2. 95 lbs, which is well above the 1. 5 lb minimum required.
Patient Sequence No: 1, Text Type: N, H10


[83279830] (b)(4) - "intervention planned in aortic re-coarctation. Six (6) years prior treatment with stent palmaz. Failure happened during crimping of the stent on bib, the teflon came loose, thus the stent was not used. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1318694-2017-00022
MDR Report Key6803422
Date Received2017-08-18
Date of Report2017-08-18
Date of Event2016-09-11
Date Mfgr Received2017-07-26
Device Manufacturer Date2015-05-05
Date Added to Maude2017-08-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICHELLE LAFLESH
Manufacturer Street2880 MAIN STREET
Manufacturer CityHOPKINTON NY 12965
Manufacturer CountryUS
Manufacturer Postal12965
Manufacturer Phone3153284491
Manufacturer G1NUMED, INC.
Manufacturer Street2880 MAIN STREET
Manufacturer CityHOPKINTON NY 12965
Manufacturer CountryUS
Manufacturer Postal Code12965
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOVERED CP STENT
Generic NameAORTIC STENT
Product CodePNF
Date Received2017-08-18
Returned To Mfg2017-08-08
Model Number427
Catalog NumberCVRDCP8Z45
Lot NumberCCP-0586
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNUMED, INC.
Manufacturer Address2880 MAIN STREET HOPKINTON NY 12965 US 12965


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.