ARCHITECT TOTAL PSA 07K70-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2017-08-18 for ARCHITECT TOTAL PSA 07K70-25 manufactured by Abbott Ireland.

Event Text Entries

[83223022] Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and accuracy testing. No adverse trend was identified for the customer's issue. Labeling was reviewed and found to be adequate. The product was not available for return. An internal panel was tested with retained kits of the likely cause reagent lot and accuracy testing met all specifications. Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
Patient Sequence No: 1, Text Type: N, H10


[83223023] The customer observed a falsely depressed total psa results on the architect i1000sr analyzer. The following data was provided for one male patient: initial 1. 52 ng / ml, repeat 1. 3 ng / ml. The sample was confirmed higher on another method (eclia) 6. 05, 5. 87 ng/ml and (immulite) 5. 45. The customer suspects interference with the patients drug proscar (finasteride). A second patient had a result of 0. 07 ng/ml, repeated on eclia as 0. 30 ng/ml, no further patient details were provided. There was no impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008344661-2017-00075
MDR Report Key6805054
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2017-08-18
Date of Report2017-08-18
Date Mfgr Received2017-07-26
Device Manufacturer Date2017-01-09
Date Added to Maude2017-08-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT IRELAND
Manufacturer StreetDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK
Manufacturer CitySLIGO NA
Manufacturer CountryEI
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT TOTAL PSA
Generic NameTOTAL PSA
Product CodeLTJ
Date Received2017-08-18
Catalog Number07K70-25
Lot Number71191FN00
Device Expiration Date2017-12-14
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT IRELAND
Manufacturer AddressDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK SLIGO NA EI NA


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.