T2X OPERATING ROOM TABLE 109682-000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-21 for T2X OPERATING ROOM TABLE 109682-000 manufactured by Imris - Deerfield Imaging, Inc..

Event Text Entries

[83531145] The imris field service engineer met with personnel from clinical engineering to inspect the or table. They concluded that the back section was not fully seated and locked into the table at the time when the incident occurred. They fully tested table movement and back section locking mechanism to ensure proper operation. All testing passed. Imris clinical education team shall communicate with (b)(6) operating room (o. R. ) personnel to ensure that they are following the instructions in the ifu and to re-educate cleaning crew and clinical staff that prepares the operating room (o. R. ) for proper usage of the or table back section. (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[83531146] At approximately 2 pm on (b)(6) 2017, the back section detached from the t2x or table with a patient during preparation for a surgery. The back section dropped to the floor and the patient slid off the back section to the floor. The patient was immediately brought to mr and imaged. No injuries were found during imaging. The following day, hospital staff confirmed that there were no injuries suffered by the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010326005-2017-00008
MDR Report Key6808034
Date Received2017-08-21
Date of Report2017-10-20
Date of Event2017-08-15
Date Mfgr Received2017-08-15
Date Added to Maude2017-08-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. SANJAY SHAH
Manufacturer Street5101 SHADY OAK ROAD
Manufacturer CityMINNETONKA MN 553434100
Manufacturer CountryUS
Manufacturer Postal553434100
Manufacturer Phone7632036380
Manufacturer G1IMRIS - DEERFIELD IMAGING, INC.
Manufacturer Street5101 SHADY OAK ROAD
Manufacturer CityMINNETONKA MN 553434100
Manufacturer CountryUS
Manufacturer Postal Code553434100
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameT2X OPERATING ROOM TABLE
Generic NameRADIOLOGIC TABLE, PRODUCT CODE: KXJ
Product CodeKXJ
Date Received2017-08-21
Model NumberT2X
Catalog Number109682-000
OperatorNURSE
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerIMRIS - DEERFIELD IMAGING, INC.
Manufacturer Address5101 SHADY OAK ROAD MINNETONKA MN 553434100 US 553434100


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.