BLOODGAS ANALYZER, ABL 800 SERIES *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2006-02-21 for BLOODGAS ANALYZER, ABL 800 SERIES * manufactured by Radiometer Medical Aps.

Event Text Entries

[403203] A sample was analyzed on the blood gas analyzer this sampler or a sample analyzed prior to this measurement introduced a clot in the blood gas analyzer. The clot was positioned in the ph chamber during measurement of sample the clot in the ph chamber led to too low ph values. For sample the ph value had not been requested. Clots may affect the sample transport and may alson lead to dilution of the sample. When a technician removed the clot the problem solved. The ph electrode was cleaned. Quality control (qc level 3+) was done, without any problem. Due to the wrong ph measurement a patient was mistreated with nabic. The patient has not suffered any degradation in state of health.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2006-00015
MDR Report Key680977
Report Source05
Date Received2006-02-21
Date of Report2006-02-06
Date of Event2006-01-05
Date Mfgr Received2006-01-05
Device Manufacturer Date2004-08-01
Date Added to Maude2006-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLISE-LOTTE RINGBY
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ 2700
Manufacturer CountryDA
Manufacturer Postal2700
Manufacturer Phone8273495
Manufacturer G1RADIOMETER MEDICAL APS
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ 2700
Manufacturer CountryDA
Manufacturer Postal Code2700
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBLOODGAS ANALYZER,
Generic NameBLOODGAS ANALYZER
Product CodeCCE
Date Received2006-02-21
Model NumberABL 800 SERIES
Catalog Number*
Lot Number754R0023N0005
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key670248
ManufacturerRADIOMETER MEDICAL APS
Manufacturer AddressAKANDEVEJ 21 * DA DK-2700
Baseline Brand NameBLOODGAS ANALYZER,
Baseline Generic NameBLOODGAS ANALYZER
Baseline Model NoABL 800 SERIES
Baseline Catalog No*
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2006-02-21