BD? DEHP FREE INFUSION SET 590502

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-22 for BD? DEHP FREE INFUSION SET 590502 manufactured by Becton Dickinson.

Event Text Entries

[83312041] Results: a sample was returned for evaluation. A hair was found inside the iv tube. In the event that new, changed, or corrected information is obtained, a supplemental report will be filed. Conclusion: based on the investigation result of the complaint case the hair inside the iv set tube was a human hair. We suppose that when the two parts(sleeve stopper and y-connector) were assembled together, a hair may have stuck onto one of them and flowed inside the y-connector due to static electricity. Defect due to insufficient cleanness of work stand or work floor or smock. Capa (b)(4) has been opened. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[83312042] It was reported that a hair was found in the tubing of a bd? Dehp free infusion set during use. No injury or medical intervention.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2243072-2017-00090
MDR Report Key6810329
Date Received2017-08-22
Date of Report2017-08-15
Date of Event2017-07-25
Date Mfgr Received2017-07-26
Device Manufacturer Date2017-04-27
Date Added to Maude2017-08-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652845
Manufacturer G1BECTON DICKINSON
Manufacturer Street1 BECTON DRIVE
Manufacturer CityFRANKLIN LAKES NJ 07417
Manufacturer CountryUS
Manufacturer Postal Code07417
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBD? DEHP FREE INFUSION SET
Generic NameIV TUBING
Product CodeFPK
Date Received2017-08-22
Returned To Mfg2017-07-27
Catalog Number590502
Lot Number2704101
OperatorNURSE
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECTON DICKINSON
Manufacturer Address1 BECTON DRIVE FRANKLIN LAKES NJ 07417 US 07417


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-08-22

© 2024 FDA.report