SOFT-TRANS EMBRYO TRANSFER CATHETER K-SOFT-5000-MO

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2017-08-22 for SOFT-TRANS EMBRYO TRANSFER CATHETER K-SOFT-5000-MO manufactured by Cook Inc.

Event Text Entries

[83318148] (b)(4). The event is currently under investigation. A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


[83318149] The cook representative visited the user facility to perform testing of unused soft-trans embryo transfer catheters in stock at the facility in an effort to find a catheter that flushes out yellow media fluid. After testing over 50 unused catheters, one catheter was finally found that the media was flushed through the catheter and came out yellow. This sample and the other devices used during the test were packaged together to return to the manufacturer for further analysis. There was no patient involvement during this testing. The issue of media fluid flushing out "yellow" was previously reported by this facility. However, samples were not provided for analysis. The two previous reports can be found in mfr. Report numbers 1820334-2017-02298 and 1820334-2017-02304.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2017-02680
MDR Report Key6811082
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2017-08-22
Date of Report2018-03-23
Date of Event2017-08-15
Date Mfgr Received2018-02-21
Device Manufacturer Date2017-05-24
Date Added to Maude2017-08-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSOFT-TRANS EMBRYO TRANSFER CATHETER
Generic NameMQF CATHETER, ASSISTED REPRODUCTION
Product CodeMQF
Date Received2017-08-22
Catalog NumberK-SOFT-5000-MO
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-22

© 2026 FDA.report