COBAS 6000 C501 MODULE 05860636001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-08-23 for COBAS 6000 C501 MODULE 05860636001 manufactured by Roche Diagnostics.

Event Text Entries

[83742317] Unique identifier (udi)#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[83742318] The customer received a questionable low alb2 albumin gen. 2 result for one patient sample. The initial result was <1. 0 g/dl and was manually released outside of the laboratory. The result was questioned by the physician and the sample was repeated with a result of 4. 8 g/dl. Information regarding any adverse event was requested but was not provided. The reagent lot number was 23581801 with an expiration date of 6/30/2018. The customer declined a service visit. There were no issues with any other samples.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01796
MDR Report Key6812759
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-08-23
Date of Report2017-09-12
Date of Event2017-08-03
Date Mfgr Received2017-08-04
Date Added to Maude2017-08-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 6000 C501 MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeCJW
Date Received2017-08-23
Model NumberC501
Catalog Number05860636001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 6000 C501 MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-08-23
Model NumberC501
Catalog Number05860636001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.