[83435085] After removing the resqpump from the box, the pump was in good condition with no signs of damage. The force gauge was zeroed when first removed from the box. When the pump was compressed for the first time, it got stuck at 50 kg. After taking the clamshell apart, the gauge mechanism visually looked to be in working order (spring and gauge in correct orientation). The pushrod stem could be pulled out but was difficult to do the first time. A black spot was identified on stem that was potentially excess loctite but it wasn't confirmed. When the pump was reassembled, the gauge worked smoothly and no other problems were detected. The accuracy of the force gauge was verified on a scale and are within specification. The investigation suggests that an excessive amount of loctite was put on the stem causing it to stick in place. This caused the force gauge to get stuck at 50 kg after being compressed.
Patient Sequence No: 1, Text Type: N, H10


[83435086] Unwrapped the resqpumps straight from the box. The force gauge on one of the pumps was stuck at 50.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003477173-2017-00005
MDR Report Key6813958
Report SourceHEALTH PROFESSIONAL
Date Received2017-08-23
Date of Report2017-08-23
Date of Event2017-07-18
Date Mfgr Received2017-07-18
Date Added to Maude2017-08-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. DIANE HOWELL
Manufacturer Street1905 COUNTY ROAD C WEST
Manufacturer CityROSEVILLE MN 55113
Manufacturer CountryUS
Manufacturer Postal55113
Manufacturer Phone6514035600
Manufacturer G1SCIENTIFIC MOLDING CORPORATION
Manufacturer Street330 SMC DRIVE
Manufacturer CitySOMERSET WI 54025
Manufacturer CountryUS
Manufacturer Postal Code54025
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRESQPUMP
Generic NameRESQPUMP
Product CodePIZ
Date Received2017-08-23
Returned To Mfg2017-07-31
Model Number12-0823-000
Lot Number51733641
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerADVANCED CIRCULATORY SYSTEMS
Manufacturer Address1905 COUNTY ROAD C WEST ROSEVILLE MN 55113 US 55113


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-23