TOTAL PROTEIN URINE/CSF GEN.3 03333825190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-08-24 for TOTAL PROTEIN URINE/CSF GEN.3 03333825190 manufactured by Roche Diagnostics.

Event Text Entries

[84084644] (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[84084645] The customer states that they received erroneous results for one patient urine sample tested for tpuc3 total protein urine/csf gen. 3 (tpuc) on a cobas 6000 c (501) module - c501. No erroneous results were reported outside of the laboratory. The sample initially resulted as 0. 09 g/l. The sample was repeated on (b)(6) 2017, resulting as 0. 19 g/l. The sample was manually diluted 1:20 and tested twice on (b)(6) 2017, resulting as 12. 82 g/l accompanied by a data flag and 12. 96 g/l accompanied by a data flag. The sample was manually diluted 1:50 and tested twice on (b)(6) 2017, resulting as 113. 38 g/l accompanied by a data flag and 112. 06 g/l accompanied by a data flag. No adverse events were alleged to have occurred with the patient. The c501 analyzer serial number was (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01808
MDR Report Key6816373
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-08-24
Date of Report2017-09-01
Date of Event2017-08-06
Date Mfgr Received2017-08-07
Date Added to Maude2017-08-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameTOTAL PROTEIN URINE/CSF GEN.3
Generic NameTURBIDIMETRIC, TOTAL PROTEIN
Product CodeJGQ
Date Received2017-08-24
Model NumberNA
Catalog Number03333825190
Lot Number223039
ID NumberNA
Device Expiration Date2018-05-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-24

© 2026 FDA.report