Event Text Entries

[83548940] This report was originally submitted on 8/10/2017, in an incorrect format. This was initially discovered on 8/23/2017, and this is being re-submitted on 8/24/2017.
Patient Sequence No: 1, Text Type: N, H10


[83548941] A complaint of an aspirated prosthesis not showing up on x-ray was reported by the patient's speech language pathologist (slp). The slp additionally reported that the "prosthesis was visualized in the patient's bronchus on a tracheal scope" and that the prosthesis "did not show up on x-ray, even with 7 views. " the x-rays were reviewed by a radiologist, attending ent, resident ent, and the slp. A full bronchoscopy was performed on the patient to remove the aspirated prosthesis and to check if a previously displaced device was aspirated. It was confirmed that the previously displaced device had been swallowed and only 1 device was aspirated. It was confirmed that the devices did not contribute to the aspiration event nor the previous event of swallowing, and that these events were related to the patient's granulation tissue. The complaintant indicated that the device was discarded and thus would not be returned for evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2025182-2017-00003
MDR Report Key6817229
Report SourceHEALTH PROFESSIONAL
Date Received2017-08-24
Date of Report2017-08-10
Date of Event2017-07-12
Date Mfgr Received2017-07-14
Device Manufacturer Date2015-11-05
Date Added to Maude2017-08-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ALYSE MENCIAS
Manufacturer Street1110 MARK AVE
Manufacturer CityCARPINTERIA CA 93013
Manufacturer CountryUS
Manufacturer Postal93013
Manufacturer G1FREUDENBERG MEDICAL
Manufacturer Street1110 MARK AVE
Manufacturer CityCARPINTERIA CA 93013
Manufacturer CountryUS
Manufacturer Postal Code93013
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBLOM-SINGER
Generic NameVOICE PROSTHESIS
Product CodeEWL
Date Received2017-08-24
Catalog NumberIN 1614-NS
Lot Number0000011020
Device Expiration Date2018-07-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFREUDENBERG MEDICAL, LLC
Manufacturer Address1110 MARK AVE CARPINTERIA CA 93013 US 93013


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-08-24