AQUASHIELD SYSTEM CO2 - FUJINON 00711553

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-24 for AQUASHIELD SYSTEM CO2 - FUJINON 00711553 manufactured by United States Endoscopy Group, Inc..

Event Text Entries

[83646215] Review of the device history record indicates no issues during manufacture; the device was manufactured to specification. The subject device was not returned to us endoscopy. The instructions for use contain the following: "prime the air/water channel prior to insertion of the endoscope into a patient. If water pressure is low, check to see that the lid cap is tightly closed. Check that the endoscope connector is firmly attached to the endoscope. " in-service training has been provided to the user.
Patient Sequence No: 1, Text Type: N, H10


[83646216] The aquashield system co2 - fujinon is intended to be used with an air or carbon dioxide (co2) source from an endoscope with the purpose of supplying sterile water to the endoscope during endoscopic procedures. It is compatible with u. S. Commercially available sterile water bottles. This device is intended to be used for not more than 24 hours. The user facility reported during a procedure the connection between the aquashield system co2 and the endoscope was leaking sterile water. There was no report of injury to the patient or the user.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1528319-2017-00025
MDR Report Key6817403
Date Received2017-08-24
Date of Report2017-08-24
Date of Event2017-08-04
Date Mfgr Received2017-08-04
Device Manufacturer Date2017-06-04
Date Added to Maude2017-08-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS COLETTA COHARA
Manufacturer Street5976 HEISLEY RD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403586251
Manufacturer G1UNITED STATES ENDOSCOPY GROUP, INC.
Manufacturer Street5976 HEISLEY RD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal Code44060
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameAQUASHIELD SYSTEM CO2 - FUJINON
Generic NameWATER BOTTLE CAP
Product CodeFEQ
Date Received2017-08-24
Model Number00711553
Catalog Number00711553
Lot Number1709622
Device Expiration Date2018-06-04
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerUNITED STATES ENDOSCOPY GROUP, INC.
Manufacturer Address5976 HEISLEY RD MENTOR OH 44060 US 44060


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-08-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.