ARCHITECT C16000 SYSTEM 03L77-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-08-25 for ARCHITECT C16000 SYSTEM 03L77-01 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[83683822] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[83683823] The customer reports falsely elevated architect clin chem phosphorus assay results being generated on an architect c16000 analyzer. A patient sample generated an initial result of 10. 1 mg/dl (customer upper normal limit is 4. 7 mg/dl). The sample retested at 3. 9 mg/dl. Controls were within specification earlier in the day. The customer discovered two bent cuvette wash nozzles and a bent r1a probe on the instrument. The customer straightened the wash nozzles and replaced the r1a probe, but failed the r1a calibration due to lls failures. A service call was initiated. No suspect results were reported from the lab. There is no impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2017-00322
MDR Report Key6819880
Report SourceHEALTH PROFESSIONAL
Date Received2017-08-25
Date of Report2017-10-09
Date Mfgr Received2017-10-09
Device Manufacturer Date2009-09-01
Date Added to Maude2017-08-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameARCHITECT C16000 SYSTEM
Generic NameAUTOMATED CHEMISTRY ANALYZER
Product CodeCEO
Date Received2017-08-25
Catalog Number03L77-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038

Device Sequence Number: 1

Brand NameARCHITECT C16000 SYSTEM
Generic NameAUTOMATED CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-08-25
Catalog Number03L77-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-25

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.