MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-08-25 for ARCHITECT C16000 SYSTEM 03L77-01 manufactured by Abbott Manufacturing Inc.
[83683822]
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10
[83683823]
The customer reports falsely elevated architect clin chem phosphorus assay results being generated on an architect c16000 analyzer. A patient sample generated an initial result of 10. 1 mg/dl (customer upper normal limit is 4. 7 mg/dl). The sample retested at 3. 9 mg/dl. Controls were within specification earlier in the day. The customer discovered two bent cuvette wash nozzles and a bent r1a probe on the instrument. The customer straightened the wash nozzles and replaced the r1a probe, but failed the r1a calibration due to lls failures. A service call was initiated. No suspect results were reported from the lab. There is no impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1628664-2017-00322 |
MDR Report Key | 6819880 |
Report Source | HEALTH PROFESSIONAL |
Date Received | 2017-08-25 |
Date of Report | 2017-10-09 |
Date Mfgr Received | 2017-10-09 |
Device Manufacturer Date | 2009-09-01 |
Date Added to Maude | 2017-08-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 224667-512 |
Manufacturer G1 | ABBOTT MANUFACTURING INC |
Manufacturer Street | 1921 HURD DRIVE |
Manufacturer City | IRVING TX 75038 |
Manufacturer Country | US |
Manufacturer Postal Code | 75038 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ARCHITECT C16000 SYSTEM |
Generic Name | AUTOMATED CHEMISTRY ANALYZER |
Product Code | CEO |
Date Received | 2017-08-25 |
Catalog Number | 03L77-01 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ABBOTT MANUFACTURING INC |
Manufacturer Address | 1921 HURD DRIVE IRVING TX 75038 US 75038 |
Brand Name | ARCHITECT C16000 SYSTEM |
Generic Name | AUTOMATED CHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2017-08-25 |
Catalog Number | 03L77-01 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MANUFACTURING INC |
Manufacturer Address | 1921 HURD DRIVE IRVING TX 75038 US 75038 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-08-25 |