STREAMLAB? ANALYTICAL WORKCELL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2017-08-25 for STREAMLAB? ANALYTICAL WORKCELL manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[83685146] Details surrounding the incident are under investigation by siemens (for example, whether the cse followed instructions in service guides). As a result of the shock, the cse powered off and unplugged the automation system to perform further troubleshooting of the air compressor. On a later date, the cse installed a new air compressor. The streamlab automation system is operating according to specifications.
Patient Sequence No: 1, Text Type: N, H10


[83685147] A siemens customer service engineer (cse) was dispatched to the customer site to perform troubleshooting on the air dryer of the streamlab automation system. The cse removed the mounting bar and received an electrical shock. The cse was not injured by the electrical shock and did not require any medical intervention or treatment. The cse powered off and unplugged the instrument to perform further troubleshooting of the air compressor due to the shock.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00667
MDR Report Key6819939
Report SourceCOMPANY REPRESENTATIVE
Date Received2017-08-25
Date of Report2017-08-25
Date of Event2017-07-27
Date Mfgr Received2017-07-27
Date Added to Maude2017-08-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactROLANDO HUALPA
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242963
Manufacturer G1INPECO S.P.A
Manufacturer StreetVIA GIVOLETTO 15 REGISTRATION #: 3005509212
Manufacturer CityVAL DELLA TORRE (TORINO) 10040
Manufacturer CountryIT
Manufacturer Postal Code10040
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTREAMLAB? ANALYTICAL WORKCELL
Generic NameSTREAMLAB? ANALYTICAL WORKCELL
Product CodeLGX
Date Received2017-08-25
Model NumberSTREAMLAB? ANALYTICAL WORKCELL
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-25

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.