BD VACUTAINER C&S TRANSFER STRAW KIT 364953

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-28 for BD VACUTAINER C&S TRANSFER STRAW KIT 364953 manufactured by Becton, Dickinson And Company.

Event Text Entries

[83852805]
Patient Sequence No: 1, Text Type: N, H10


[83852806] Staff have reported difficulty transferring urine from specimen cups (in patients with no foley) to the bd vacutainer tube using the c&s transfer straw kit. This has resulted to exposures and near misses with urine being spilled, which is particularly concerning when dealing with patients receiving chemotherapy. The straw end of the kit has to be held in the specimen cup while the vacutainer tube is pushed down onto the needle on the other end of the straw kit. Staffs have reported difficulty in getting the needle to penetrate the yellow rubber cap on the end of the vacutainer, resulting in jarring of the specimen cup and spillage of urine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6823261
MDR Report Key6823261
Date Received2017-08-28
Date of Report2017-08-10
Date of Event2017-08-05
Report Date2017-08-10
Date Reported to FDA2017-08-10
Date Reported to Mfgr2017-08-10
Date Added to Maude2017-08-28
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBD VACUTAINER C&S TRANSFER STRAW KIT
Generic NameURINE COLLECTION DEVICE
Product CodeOIE
Date Received2017-08-28
Catalog Number364953
Lot Number7097784
OperatorNURSE
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBECTON, DICKINSON AND COMPANY
Manufacturer Address1 BECTON DR. FRANKLIN LAKES NJ 07417 US 07417


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-28

© 2026 FDA.report