TRIGL TRIGLYCERIDES 05171407190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-08-28 for TRIGL TRIGLYCERIDES 05171407190 manufactured by Roche Diagnostics.

Event Text Entries

[84000275] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[84000276] The customer stated that they received erroneous results for two patient samples tested for triglycerides (trig) on a cobas 8000 c 702 module (c702). On (b)(6) 2017, the customer stated that both samples were tested within the past month, but could not provide details on exactly when they were tested. The date of event is not known. Both samples initially resulted with values that were approximately 240 mg/dl and these results were reported outside of the laboratory. The samples were diluted and repeated, with each resulting as approximately 4000 mg/dl. The patients were not adversely affected. The c702 analyzer serial (b)(4). When the customer's trig application settings were checked in the system software, it was determined that they were not using the current prozone limit settings for the trig application. The customer's trig prozone limit settings were updated and the test was re-calibrated.
Patient Sequence No: 1, Text Type: D, B5


[119832446] The customer has had no further issues since updating the application settings. The event was caused by user error as the customer did not update their application setting per product labeling.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2017-01835
MDR Report Key6824858
Date Received2017-08-28
Date of Report2017-09-20
Date of Event2017-07-09
Date Mfgr Received2017-08-09
Date Added to Maude2017-08-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameTRIGL TRIGLYCERIDES
Generic NameTRIGLYCERIDES GPO WITHOUT FREE GLYCEROL
Product CodeCDT
Date Received2017-08-28
Model NumberNA
Catalog Number05171407190
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-28

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.