EP-WORKMATE? SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 120 CHANNEL 09-1463-0120

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-08-29 for EP-WORKMATE? SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 120 CHANNEL 09-1463-0120 manufactured by St. Jude Medical, Inc..

Event Text Entries

[83869939] When the amplifier was booting it gave only one tone and would not communicate with the workstation. The procedure was abandoned and there were no adverse consequences to the patient.
Patient Sequence No: 1, Text Type: D, B5


[108386893] Based on the information provided to abbott and the investigation performed, the root cause of the reported event was isolated to abnormal functionality of the single board computer pca. The device history record was reviewed to ensure that each manufacturing and inspection operation was performed. The review determined the process was performed and completed in accordance with abbott specifications and procedures.
Patient Sequence No: 1, Text Type: N, H10


[108386897] When the amplifier was booting up it gave only one tone and would not communicate with the workstation. The paroxysmal supraventricular tachycardia (psvt) ablation procedure was cancelled and there were no adverse consequences to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2184149-2017-00024
MDR Report Key6826991
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-08-29
Date of Report2017-10-03
Date of Event2017-08-15
Date Mfgr Received2017-09-22
Device Manufacturer Date2007-11-02
Date Added to Maude2017-08-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDENISE JOHNSON
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517564470
Manufacturer G1ST. JUDE MEDICAL, INC.
Manufacturer StreetONE ST. JUDE MEDICAL DRIVE
Manufacturer CityST. PAUL MN 55117
Manufacturer CountryUS
Manufacturer Postal Code55117
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEP-WORKMATE? SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 120 CHANNEL
Generic NameEXTERNAL PROGRAMMABLE PACEMAKER PULSE GENERATOR
Product CodeJOQ
Date Received2017-08-29
Model Number09-1463-0120
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, INC.
Manufacturer AddressONE ST. JUDE MEDICAL DRIVE ST. PAUL MN 55117 US 55117


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.