TRIAGE PROFILER SOB PANEL 25T 97300EU

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,foreign,health profe report with the FDA on 2017-08-29 for TRIAGE PROFILER SOB PANEL 25T 97300EU manufactured by Alere San Diego, Inc..

Event Text Entries

[83987812] Investigation conclusion: the customer's complaint was not replicated during in-house testing with retains of device lot w62754b. No issues with ckmb recovery were observed. Manufacturing batch records for the lot were reviewed and found that the lot met release specifications. Based on the information available, there is no indication of a product deficiency and no corrective action is required. This event is being reported as device is same/similar to 510k: k080269.
Patient Sequence No: 1, Text Type: N, H10


[83987813] On (b)(6) 2017, a nurse tested her own edta blood sample using the triage profiler sob panel and received a normal ck-mb result of <1. 0 ng/ml. The sample was sent to the laboratory and produced an abnormal ck-mb result of 37 u/l. No treatment or intervention was administered or withheld based on the triage ck-mb result. Facility triage ck-mb normal cutoff= 4. 3ng/ml. Facility laboratory ck-mb normal cutoff= 24u/l.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2017-00138
MDR Report Key6828989
Report SourceCONSUMER,FOREIGN,HEALTH PROFE
Date Received2017-08-29
Date of Report2017-08-22
Date of Event2017-08-22
Date Mfgr Received2017-08-22
Date Added to Maude2017-08-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE PROFILER SOB PANEL 25T
Generic NameCARDIOPROFILER PANEL
Product CodeJHX
Date Received2017-08-29
Model Number97300EU
Lot NumberW62754B
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-29

© 2026 FDA.report