MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-08-31 for COBAS 6000 E 601 MODULE E601 04745922001 manufactured by Roche Diagnostics.
[86008815]
(b)(4).
Patient Sequence No: 1, Text Type: N, H10
[86008816]
The customer obtained questionable (false-negative) results for three patient samples using the elecsys hsv-1 igg immunoassay (hsv-1) on the cobas 6000 e 601 module. All initial results were released outside the laboratory. No data flags or alarms occurred. Refer to the attachment to this medwatch for all patient data. No adverse events occurred. The hsv-1 reagent lot number is 21450900 with an expiration date of 04/30/2018. The customer stated the labcorp method was for research only or a lab-developed method. The customer has stopped reporting hsv-1 results but continues to run samples and send aliquots out of the laboratory for correlation. The field service representative could not find a cause for the issue. He ran performance and voltage checks which passed. The customer stated that there was no sample left for investigation. Investigation activities are ongoing.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2017-01876 |
MDR Report Key | 6835079 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2017-08-31 |
Date of Report | 2017-10-09 |
Date of Event | 2017-08-01 |
Date Mfgr Received | 2017-08-14 |
Date Added to Maude | 2017-08-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | HITACHI HIGH TECH CORP. |
Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
Manufacturer City | IBARAKI 312-8504 |
Manufacturer Country | JA |
Manufacturer Postal Code | 312-8504 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS 6000 E 601 MODULE |
Generic Name | IMMUNOCHEMISTRY ANALYZER |
Product Code | MXJ |
Date Received | 2017-08-31 |
Model Number | E601 |
Catalog Number | 04745922001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Brand Name | COBAS 6000 E 601 MODULE |
Generic Name | IMMUNOCHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2017-08-31 |
Model Number | E601 |
Catalog Number | 04745922001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-08-31 |