COBAS 6000 E 601 MODULE E601 04745922001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-08-31 for COBAS 6000 E 601 MODULE E601 04745922001 manufactured by Roche Diagnostics.

Event Text Entries

[86008815] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[86008816] The customer obtained questionable (false-negative) results for three patient samples using the elecsys hsv-1 igg immunoassay (hsv-1) on the cobas 6000 e 601 module. All initial results were released outside the laboratory. No data flags or alarms occurred. Refer to the attachment to this medwatch for all patient data. No adverse events occurred. The hsv-1 reagent lot number is 21450900 with an expiration date of 04/30/2018. The customer stated the labcorp method was for research only or a lab-developed method. The customer has stopped reporting hsv-1 results but continues to run samples and send aliquots out of the laboratory for correlation. The field service representative could not find a cause for the issue. He ran performance and voltage checks which passed. The customer stated that there was no sample left for investigation. Investigation activities are ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01876
MDR Report Key6835079
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-08-31
Date of Report2017-10-09
Date of Event2017-08-01
Date Mfgr Received2017-08-14
Date Added to Maude2017-08-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 6000 E 601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeMXJ
Date Received2017-08-31
Model NumberE601
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 6000 E 601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-08-31
Model NumberE601
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-08-31