MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-08-31 for 1.RT-5100 manufactured by Nidek Co., Ltd..
[84997907]
No serious injury occurred and no known impact or consequence to the doctor involved. However, nidek incorporated considers this issue a reportable event as the device had malfunctioned and has a potential to cause or contribute to a serious injury if the malfunction were to recur. Nidek inc. Is aware of this issue and corrective action has been implemented under recall z-1245-2016. Nidek hired a third party ((b)(4)) to perform the correction as per z-1245-2016.
Patient Sequence No: 1, Text Type: N, H10
[84997908]
On august 18, 2017, nidek inc. Distributor sales manager received an email from the director of customer care from (b)(6) that a customer (whom is a doctor) was upset that the near point chart rod on their rt-5100 fell several times and had hit him on the head on multiple occasions. The doctor has requested a service repair. The complainant had confirmed that no bruise or no serious injury occurred, and thereby, medical treatment/intervention was unnecessary.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 0002936921-2017-00021 |
MDR Report Key | 6835428 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2017-08-31 |
Date of Report | 2017-08-31 |
Date of Event | 2017-08-18 |
Date Mfgr Received | 2017-08-18 |
Device Manufacturer Date | 2014-04-02 |
Date Added to Maude | 2017-08-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS. PREETI GANDHI BHATIA |
Manufacturer Street | 47651 WESTINGHOUSE DRIVE |
Manufacturer City | FREMONT CA 945397474 |
Manufacturer Country | US |
Manufacturer Postal | 945397474 |
Manufacturer Phone | 8002239044 |
Manufacturer G1 | NIDEK CO., LTD. |
Manufacturer Street | 34-14 HIROISHI |
Manufacturer City | GAMAGORI, AICHI 443-0038 |
Manufacturer Country | JA |
Manufacturer Postal Code | 443-0038 |
Single Use | 3 |
Remedial Action | RC |
Previous Use Code | 3 |
Removal Correction Number | Z-1245-2016 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | 1.RT-5100 |
Generic Name | REFRACTOR |
Product Code | HKN |
Date Received | 2017-08-31 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | NIDEK CO., LTD. |
Manufacturer Address | 34-14 HIROISHI GAMAGORI, AICHI 443-0038 JA 443-0038 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-08-31 |