ELECSYS CK-MB STAT IMMUNOASSAY 05894808190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-09-01 for ELECSYS CK-MB STAT IMMUNOASSAY 05894808190 manufactured by Roche Diagnostics.

Event Text Entries

[85776891] Unique identifier (udi)#: (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[85776892] The customer noticed a shift in elecsys ck-mb immunoassay results for qc and patient samples when putting reagent lot 222123 into use. The customer performed a patient comparison and of the data provided for 13 patient samples, only the results for two patient samples were discrepant. Patient 1 original result with an unknown lot number was 22. With reagent lot 222123, the result was 27. With reagent lot 164858, the result was 22. Patient 2 original result with an unknown lot number was 130. With reagent lot 222123, the result was 150. With reagent lot 164858, the result was 120. The unit of measure was requested but was not provided. Information concerning if any erroneous result was reported outside of the laboratory or if any patient was adversely affected was requested, but it was not provided. The customer used cobas 6000 e 601 module serial number (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01883
MDR Report Key6837422
Date Received2017-09-01
Date of Report2017-11-06
Date of Event2017-08-08
Date Mfgr Received2017-08-08
Date Added to Maude2017-09-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameELECSYS CK-MB STAT IMMUNOASSAY
Generic NameCOLORIMETRIC METHOD, CPK OR ISOENZYMES
Product CodeJHY
Date Received2017-09-01
Model NumberNA
Catalog Number05894808190
Lot Number222123
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.