ENDOGATOR AUXILIARY WATER JET CONNECTOR 100310

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-09-01 for ENDOGATOR AUXILIARY WATER JET CONNECTOR 100310 manufactured by Medivators.

Event Text Entries

[84610964] A facility reported that blood/fluid backflow was observed in the irrigation pump tubing during a procedure while using medivators endogator auxiliary water jet connector. The initial report included photographic evidence of the tubing and connector involved in this incident. It is unknown if the facility replaced/switched out both the tubeset and auxilliary connector after the procedure as instructed in medivators endogator endoscopy irrigation tubing ifu. There is potential cross contamination between patient procedures. Medivators was unable to confirm if the facility was using the endogator auxiliary water jet connector with medivators endogator tubing which it is specifically intended for use with, thus potential off label use of the device. To date, there have been no reported illnesses or injuries as a result of this incident. This complaint will continue to be monitored within medivators complaint system.
Patient Sequence No: 1, Text Type: N, H10


[84610965] A facility reported blood/fluid backflow in the tubing to the pump during a procedure using medivators auxiliary water jet connector. There is potential for patient cross contamination.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1651395-2017-00002
MDR Report Key6839309
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-09-01
Date of Report2017-09-01
Date of Event2017-08-02
Date Mfgr Received2017-08-02
Device Manufacturer Date2017-04-25
Date Added to Maude2017-09-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactALEX NELSON
Manufacturer Street14605 28TH AVE N
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635094799
Manufacturer G1MEDIVATORS
Manufacturer Street3150 POLLOK DRIVE
Manufacturer CityCONROE TX 77303
Manufacturer CountryUS
Manufacturer Postal Code77303
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENDOGATOR AUXILIARY WATER JET CONNECTOR
Generic NameAUXILIARY WATER JET CONNECTOR
Product CodeFEQ
Date Received2017-09-01
Returned To Mfg2017-08-22
Model Number100310
Lot Number855261
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDIVATORS
Manufacturer Address3150 POLLOK DRIVE CONROE TX 77303 US 77303


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-09-01

© 2026 FDA.report