TRIAGE D-DIMER PANEL 98100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional, report with the FDA on 2017-09-01 for TRIAGE D-DIMER PANEL 98100 manufactured by Alere San Diego, Inc..

Event Text Entries

[86574165] Investigation conclusion: customer's complaint was not replicated with in-house testing of retain lot w62173b. No issues with analyte recovery were observed and the product performed as expected. Manufacturing batch records for the lot were reviewed and found that the lot met release specifications. Based on the information available, there is no indication of a product deficiency and no corrective action is required.
Patient Sequence No: 1, Text Type: N, H10


[86574166] The customer reported performing a d-dimer method comparison using the triage system and the acl elite between (b)(6) 2016 and (b)(6) 2017. On (b)(6) 2017, the customer reported that the one of the physicians at the facility was concerned with the results from the method comparison. All results were obtained using patient samples. Dates and results are included below: sample number - date - triage d-dimer result - acl elite d-dimer result:(b)(6). The customer was unable to provide the normal d-dimer cutoff used for either test. The triage d-dimer panel had not been implemented at the facility for patient care.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2017-00140
MDR Report Key6839535
Report SourceCONSUMER,HEALTH PROFESSIONAL,
Date Received2017-09-01
Date of Report2017-08-15
Date Mfgr Received2017-08-15
Date Added to Maude2017-09-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE D-DIMER PANEL
Generic NameFIBRIN SPLIT PRODUCTS
Product CodeGHH
Date Received2017-09-01
Model Number98100
Lot NumberW62173B
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.