VOLDYNE 2500 VOLUMETRIC EXERCISER 8884719025

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-09-05 for VOLDYNE 2500 VOLUMETRIC EXERCISER 8884719025 manufactured by Teleflex Medical.

Event Text Entries

[85109551] (b)(4). A visual, dimensional, and functional inspection of the device involved in this complaint could not be conducted since the device was not returned to the manufacturer at the time of this report. A device history record investigation did not show issues related to this complaint. Review of revision of (b)(4) rev 06 was performed and the failure mode is already included. No update is required. A sample of 13 devices were taken from the current production. The samples were functionally inspected. The issue reported "failure to function" was not observed. Customer complaint cannot be confirmed based on the information available at this time. No corrective actions can be assigned. If the device sample becomes available at a later date, this report will be updated accordingly.
Patient Sequence No: 1, Text Type: N, H10


[85109552] Customer complaint alleges that " at the patient's breathing exercise, the flow cup moved up but when falling down the cup got stuck". No report of patient harm or injury. No report of necessary medical intervention. Patient condition reported as "fine".
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003898360-2017-00967
MDR Report Key6843109
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-09-05
Date of Report2017-08-30
Date of Event2017-08-23
Date Mfgr Received2018-02-06
Device Manufacturer Date2017-01-31
Date Added to Maude2017-09-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetRANCHO EL DESCANSO
Manufacturer CityTECATE 21478
Manufacturer CountryMX
Manufacturer Postal Code21478
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVOLDYNE 2500 VOLUMETRIC EXERCISER
Generic NameSPIROMETER, THERAPEUTIC (INCEN
Product CodeBWF
Date Received2017-09-05
Returned To Mfg2017-10-30
Catalog Number8884719025
Lot Number73A1700767
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-05

© 2026 FDA.report