METAL BUCKLE GAIT BELT, STANDARD WHITE 6524

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-09-06 for METAL BUCKLE GAIT BELT, STANDARD WHITE 6524 manufactured by Posey Products Llc.

Event Text Entries

[85377211] This report is based solely on a product review and no incidents were reported. If the buckle(s) pre-releases or is backwards, the complaint is reportable. These conditions may allow the buckle to initially close but it is possible that the buckle could later unlatch while in use. The instructions for use were reviewed and found to provide adequate instructions and warnings for safe and effective use of the device. The application instructions caution the user to always inspect before each use: check for broken stitches or parts; torn, cut or frayed material; or buckles that are cracked or broken and do not hold securely. Never use soiled or damaged products. At this time there is no evidence that a manufacturing nonconformity contributed to the reported complaint, and the instructions for use were reviewed and determined to provide adequate instructions and warnings for the safe and effective use of the device. Therefore, no corrective or preventative actions are necessary at this time. (b)(4). Device not available. Customer review.
Patient Sequence No: 1, Text Type: N, H10


[85377212] Post market surveillance search on (b)(6) found several reports about the posey gait belts: buckle was sewn on wrong, the quick release buckle does not hold when tightened. The date these issue were discovered is unknown and no incidents or injuries were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020362-2017-00064
MDR Report Key6845380
Date Received2017-09-06
Date of Report2017-08-14
Date Mfgr Received2017-08-14
Date Added to Maude2017-09-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactWILLIAM HINCY
Manufacturer StreetPOSEY PRODUCTS LLC 5635 PECK ROAD
Manufacturer CityARCADIA CA 91006
Manufacturer CountryUS
Manufacturer Postal91006
Manufacturer Phone6264433143
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameMETAL BUCKLE GAIT BELT, STANDARD WHITE
Generic NameAID, TRANSFER
Product CodeIKX
Date Received2017-09-06
Model Number6524
Catalog Number6524
Lot NumberUNK
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPOSEY PRODUCTS LLC
Manufacturer Address5635 PECK ROAD ARCADIA CA 91006 US 91006


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-06