[86104390]
A customer from (b)(6) reported to biom? Rieux out of range low results when testing the internal quality control virotrol from bio-rad in association with vidas? Cmv igm (lot 1005685430), when compared to the specification the customer put in place, from results based on previous lots. Calibration values: compartment b1 - s1: bdf = 148/1125 rfv [714 - 1548], compartment b2 - s1: bdf = 144/1050 rfv, compartment b3 - c1: bdf = 144/1614 rfv vt = 1. 48 [1. 12 - 1. 60], compartment b4 - c2: bdf = 144 / -6 rfv vt = 0 [ equivocal (not conformed), (b)(6) 2017: bdf = 146/863 rfv / vt = 0. 79 => equivocal (non-compliant), (b)(6) 2017: bdf = 150/802 rfv / vt = 0. 73 => equivocal (not conformed), (b)(6) 2017: bdf = 157/920 rfv / vt = 0. 82 => equivocal (non-compliant). Recalibration with vortexing for 20 seconds and initiation of icq after homogenization by 20 turns resulted in the ciq as always equivocal. Compartment b1 - s1: bdf = 149/1155 rfv, compartment b2 - s1: bdf = 150/1269 rfv, compartment b3 - c1: bdf = 147/1596 rfv vt = 1. 31, compartment b4 - c2: bdf = 146 / -9 rfv / vt = -0. 00, compartment b5 virotrol: bdf = 144/970 rfv / vt = 0. 80 => equivocal. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5