MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-09-08 for VIDAS? CMV IGM 30205 manufactured by Biomerieux, Sa.
[86843159]
A customer in (b)(6) notified biom? Rieux of low out of range results when testing the internal quality control biorad torch with vidas? Cmv igm, lot 1005756380. The value obtained was 0. 97 for set range of 1. 188 / 1. 658. Calibration was within the range, but in the lower boundary for c1: s1: 1077/1179 rfv, c1: 1312 rfv / vt=1. 19 ( range 1. 11-1. 59 vt). Calibration was repeated twice, and the control c1 was higher but the ciq was still too low: s1 1024:1074, c1 1538 rfv / vt= 1. 46, cqi 1. 00 too low. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. There was no patient associated with this discrepant quality control result. A biom? Rieux internal investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3002769706-2017-00277 |
MDR Report Key | 6852350 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2017-09-08 |
Date of Report | 2018-07-17 |
Date Mfgr Received | 2018-06-20 |
Device Manufacturer Date | 2017-06-09 |
Date Added to Maude | 2017-09-08 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS. ELLEN WELTMER |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal | 63042 |
Manufacturer Phone | 3147317301 |
Manufacturer G1 | BIOMERIEUX, SA |
Manufacturer Street | 376 CHEMIN DE L'ORME |
Manufacturer City | MARCY L'ETOILE, 69280 |
Manufacturer Country | FR |
Manufacturer Postal Code | 69280 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | VIDAS? CMV IGM |
Generic Name | VIDAS? CMV IGM |
Product Code | LKQ |
Date Received | 2017-09-08 |
Catalog Number | 30205 |
Lot Number | 1005756380 |
Device Expiration Date | 2018-04-13 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMERIEUX, SA |
Manufacturer Address | 376 CHEMIN DE L'ORME MARCY L'ETOILE, 69280 FR 69280 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-09-08 |