VIDAS? CMV IGM 30205

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-09-08 for VIDAS? CMV IGM 30205 manufactured by Biomerieux, Sa.

Event Text Entries

[86843159] A customer in (b)(6) notified biom? Rieux of low out of range results when testing the internal quality control biorad torch with vidas? Cmv igm, lot 1005756380. The value obtained was 0. 97 for set range of 1. 188 / 1. 658. Calibration was within the range, but in the lower boundary for c1: s1: 1077/1179 rfv, c1: 1312 rfv / vt=1. 19 ( range 1. 11-1. 59 vt). Calibration was repeated twice, and the control c1 was higher but the ciq was still too low: s1 1024:1074, c1 1538 rfv / vt= 1. 46, cqi 1. 00 too low. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. There was no patient associated with this discrepant quality control result. A biom? Rieux internal investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00277
MDR Report Key6852350
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-09-08
Date of Report2018-07-17
Date Mfgr Received2018-06-20
Device Manufacturer Date2017-06-09
Date Added to Maude2017-09-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street376 CHEMIN DE L'ORME
Manufacturer CityMARCY L'ETOILE, 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? CMV IGM
Generic NameVIDAS? CMV IGM
Product CodeLKQ
Date Received2017-09-08
Catalog Number30205
Lot Number1005756380
Device Expiration Date2018-04-13
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address376 CHEMIN DE L'ORME MARCY L'ETOILE, 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.