I-STAT CREA CARTRIDGE 03P84-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-09-11 for I-STAT CREA CARTRIDGE 03P84-25 manufactured by Abbott Point Of Care.

Event Text Entries

[86701144] Apoc incident # (b)(4). Apoc labeling will be evaluated during the investigation as pertaining to the event.
Patient Sequence No: 1, Text Type: N, H10


[86701145] On (b)(6) 2017, abbott point of care was contacted by a customer regarding i-stat crea cartridges that yielded a suspected discrepant crea result on a patient. There was no additional patient information at the time of this report. There was no repeat on i-stat. Return product is not available for investigation. (b)(6). There are no injuries associated with this event. The investigation is underway.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2245578-2017-00287
MDR Report Key6854268
Report SourceHEALTH PROFESSIONAL
Date Received2017-09-11
Date of Report2017-10-23
Date of Event2017-08-13
Date Mfgr Received2017-10-19
Device Manufacturer Date2017-05-13
Date Added to Maude2017-09-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLINDA MACZUSZENKO
Manufacturer Street400 COLLEGE ROAD
Manufacturer CityPRINCETON NJ 08540
Manufacturer CountryUS
Manufacturer Postal08540
Manufacturer Phone6136885949
Manufacturer G1ABBOTT POINT OF CARE CANADA LTD.
Manufacturer Street185 CORKSTOWN ROAD
Manufacturer CityOTTAWA, ONTARIO K2H 8V4
Manufacturer CountryCA
Manufacturer Postal CodeK2H 8V4
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameI-STAT CREA CARTRIDGE
Generic NameCREA CARTRIDGE
Product CodeCGL
Date Received2017-09-11
Model NumberNA
Catalog Number03P84-25
Lot NumberA17133
Device Expiration Date2017-10-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT POINT OF CARE
Manufacturer Address400 COLLEGE ROAD PRINCETON NJ 085406607 US 085406607


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2017-09-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.