BIOMET OFFSET TIBIAL TRAY ADAPTER N/A 141491

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,health profession report with the FDA on 2017-09-14 for BIOMET OFFSET TIBIAL TRAY ADAPTER N/A 141491 manufactured by Zimmer Biomet, Inc..

Event Text Entries

[86311212] (b)(4). Udi: (b)(4). Concomitant medical product: biomet splined knee stem, cat#: 141618 lot#: 169810. Biomet offset tibial tray, cat#: 141484 lot#: 212390. Customer has indicated that the product is in process of being returned to zimmer biomet for investigation. Once the investigation has been completed, a follow-up mdr will be submitted. If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly. Zimmer biomet will continue to monitor for trends.
Patient Sequence No: 1, Text Type: N, H10


[86311213] It was reported that the offset adapter was not fully seated. It was noted that the implants were not assembled correctly. The screw was removed and the offset adapter was then disassembled from the tibial tray, at which time the offset adapter was found to be damaged. A replacement part was unavailable, so the surgeon re-trialed using a different size. The construct would not fully seat so the surgeon removed the polyethylene plug from the augment holes. The implants then seated and the case was completed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001825034-2017-07109
MDR Report Key6866733
Report SourceDISTRIBUTOR,HEALTH PROFESSION
Date Received2017-09-14
Date of Report2018-09-27
Date of Event2017-08-23
Date Mfgr Received2018-09-14
Device Manufacturer Date2012-06-15
Date Added to Maude2017-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CHRISTINA ARNT
Manufacturer Street56 E. BELL DR.
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5745273773
Manufacturer G1ZIMMER BIOMET, INC.
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameBIOMET OFFSET TIBIAL TRAY ADAPTER
Generic NamePROSTHESIS, KNEE
Product CodeMBV
Date Received2017-09-14
Returned To Mfg2017-09-26
Model NumberN/A
Catalog Number141491
Lot Number149400
ID NumberSEE H10 NARRATIVE
OperatorPHYSICIAN
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZIMMER BIOMET, INC.
Manufacturer Address56 E. BELL DRIVE WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.