EVA, COMBINED MACHINE INCLUDING LASER (DORC CONNECTOR) 8000.COM05 N/A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2017-09-14 for EVA, COMBINED MACHINE INCLUDING LASER (DORC CONNECTOR) 8000.COM05 N/A manufactured by Dutch Ophthalmic Research Center.

Event Text Entries

[86244374] Any additional information received will be provided in a supplemental report.
Patient Sequence No: 1, Text Type: N, H10


[86244375] During surgery, the screen froze and therefore the eva was turned off and on. In the meantime, the eye turned to hypotonic. The patient got a hypotonic eye. The surgeon intervened manually. According to (b)(4) product specialist, the surgeon had to stabilize the eye and its pressure before rebooting the eva, this would have prevented the hypotonic eye. The frozen screen will be investigated once the screen is replaced and returned for investigation. Any further information will be reported in a supplemental report. Please note that this incident occurred in (b)(6).
Patient Sequence No: 1, Text Type: D, B5


[117916115] The affected screen has been received and the investigation has started. The results of the investigation will be reported in a supplemental report.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1222074-2017-00010
MDR Report Key6866886
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2017-09-14
Date of Report2018-08-23
Date of Event2017-08-08
Date Mfgr Received2018-07-24
Device Manufacturer Date2016-11-09
Date Added to Maude2017-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ANNELIES POTH
Manufacturer StreetSCHEIJDELVEQEG 2 BUILDING 1
Manufacturer CityZUILAND, NETHERLANDS NH 3214VN
Manufacturer CountryNL
Manufacturer Postal3214 VN
Manufacturer Phone6037786929
Manufacturer G1DUTCH OPHTHALMIC RESEARCH CENTER
Manufacturer StreetSCHEIJDELVEWEG 2 3214 VN
Manufacturer CityZUIDLAND, 3214VN
Manufacturer CountryNL
Manufacturer Postal Code3214 VN
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEVA, COMBINED MACHINE INCLUDING LASER (DORC CONNECTOR)
Generic NamePHACOEMULSIFICATION/VITRECTOMY SYSTEM
Product CodeMLZ
Date Received2017-09-14
Returned To Mfg2017-11-23
Model Number8000.COM05
Catalog NumberN/A
Lot NumberN/A
OperatorPHYSICIAN
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDUTCH OPHTHALMIC RESEARCH CENTER
Manufacturer AddressSCHEIJDELVEWEG 2 3214 VN ZUIDLAND, 3214VN NL 3214 VN


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2017-09-14