COBAS 6000 C (501) MODULE C501 04745914001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-09-15 for COBAS 6000 C (501) MODULE C501 04745914001 manufactured by Roche Diagnostics.

Event Text Entries

[86416929] (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[86416930] The customer stated that they received erroneous results for one patient sample tested for dig digoxin (dig) on a cobas 6000 c (501) module - c501. No erroneous results were reported outside of the laboratory. The sample initially resulted as 1. 17 ng/ml. The sample was repeated, resulting as 0. 97 ng/ml. The sample was repeated three more times on (b)(6) 2017, resulting as 1. 44 ng/ml, 1. 25 ng/ml, and 1. 03 ng/ml. No adverse events were alleged to have occurred with the patient. It was recommended to the customer to collect another sample from the patient. The dig reagent lot number was 227984. The reagent expiration date was asked for, but not provided. The customer performed precision studies with another patient sample. Quality control recovery was stated to be "not so good".
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-02003
MDR Report Key6868799
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-09-15
Date of Report2017-10-09
Date of Event2017-08-25
Date Mfgr Received2017-08-28
Date Added to Maude2017-09-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 6000 C (501) MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeKXT
Date Received2017-09-15
Model NumberC501
Catalog Number04745914001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 6000 C (501) MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-09-15
Model NumberC501
Catalog Number04745914001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-15